Development of Novel and Highly Specific ssDNA-Aptamer-Based Electrochemical Biosensor for Rapid Detection of Mercury (II) and Lead (II) Ions in Water

CHEMOSENSORS(2019)

引用 47|浏览10
暂无评分
摘要
In this work, we report on the development of an electrochemical biosensor for high selectivity and rapid detection of Hg2+ and Pb2+ ions using DNA-based specific aptamer probes labeled with ferrocene (or methylene blue) and thiol groups at their 5 ' and 3 ' termini, respectively. Aptamers were immobilized onto the surface of screen-printed gold electrodes via the SH (thiol) groups, and then cyclic voltammetry and impedance spectra measurements were performed in buffer solutions with the addition of HgCl2 and PbCl2 salts at different concentrations. Changes in 3D conformation of aptamers, caused by binding their respective targets, e.g., Hg2+ and Pb2+ ions, were accompanied by an increase in the electron transfer between the redox label and the electrode. Accordingly, the presence of the above ions can be detected electrochemically. The detection of Hg2+ and Pb2+ ions in a wide range of concentrations as low as 0.1 ng/mL (or 0.1 ppb) was achieved. The study of the kinetics of aptamer/heavy metal ions binding gave the values of the affinity constants of approximately 9.10(-7) mol, which proved the high specificity of the aptamers used.
更多
查看译文
关键词
aptamer sensors,electrochemical sensors,heavy metal ions,cyclic voltammetry,impedance spectroscopy,binding kinetics