Implantation of the Micra transcatheter pacing system: A single center North India experience.

Indian pacing and electrophysiology journal(2020)

引用 1|浏览2
暂无评分
摘要
s: The device was successfully implanted in 28 subjects. The mean intraoperative sensing value was 9.04 ± 1.5 mV and the impedance was 766.89 ± 213.9 Ω. At discharge from hospital, those values were 13.2 ± 15.83 mV and 855 ± 111.7, respectively. The recommended pacing threshold value as achieved in all subjects was 0.78 V, i.e. ≤ 1 V at 0.24 ms. There was no adverse event or complications reported for any of the subjects. Mean time from hospitalization to discharge was 1.5 days. Implantation of leadless pacemakers is feasible, safe and provides advantages over the conventional system.
更多
查看译文
关键词
Bradycardia,Leadless pacemaker,Leadless pacing system, cardiac pacing,Micra TPS,Micra transcatheter pacing system