Albuminuria detection using graphene oxide-mediated fluorescence quenching aptasensor.

MethodsX(2020)

引用 13|浏览8
暂无评分
摘要
A simple and sensitive graphene oxide-mediated fluorescence quenching aptasensor is developed to quantify albuminuria in urine samples. The developed aptasensor used the specific target binding property of aptamer and fluorescence quenching property of graphene oxide to determine the concentration of human serum albumin in urine. The limit of detection of the developed platform is 0.05 µg.mL−1 and the detection range is 0.1–600 µg.mL−1, which covers the albuminuria concentration range present in normal human urine and the urine of the patient with chronic kidney disease. This approach can be modified to measure albuminuria using a high-throughput quantification platform and portable point of care testing. In addition, the production cost for one reaction is cheaper than those for the standard automated method. Therefore, this aptasensor has significant potential for commercialization and public use.•Our protocol is customized by using the fluorescence quenching property of graphene oxide and specific binding property of human serum albumin aptamer to detect human serum albumin in urine sample•The limit of detection of our developed platform is 0.05 µg.mL−1•The detection range of our aptasensor is 0.1–600 µg.mL−1
更多
查看译文
关键词
Albuminuria detection by graphene oxide-mediated fluorescence quenching aptasensor.
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要