Motor potential evoked by transcranial magnetic stimulation depends on the placement protocol of recording electrodes: a pilot study

BIOMEDICAL PHYSICS & ENGINEERING EXPRESS(2020)

引用 3|浏览16
暂无评分
摘要
Objective: There seems to be no consensus in the literature regarding the protocol of surface electromyography (sEMG) electrode placement for recording motor evoked potentials (MEP) in transcranial magnetic stimulation (TMS) applications. Thus, the aim of this study was to investigate the effect on the MEP amplitude bytwo different protocols for electrode placement. Methods: sEMG electrodes were placed on three upper arm muscles (biceps brachii,flexor carpi radialis,andflexor pollicis brevis) of six right-handed subjects following two different protocols (1 and 2), which varied according to the interelectrode distance and location relative to the muscle. TMS pulses were applied to thehotspotofbiceps brachii, while sEMGwas recorded from the two protocols and for each muscle simultaneously. Main Results: Greater MEP amplitudes were obtained for Protocol 1 compared to Protocol 2 (P < 0.05). Significance: Different electrode placement protocols may result in distinct MEP amplitudes, which should be taken into account when adjusting the intensity on single and repetitive TMS sessions.
更多
查看译文
关键词
motor evoked potential,surface electromyography,corticospinal excitability,transcranial magnetic stimulation,TMS,MEP
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要