Characterisation and mapping of scattered radiation fields in interventional radiology theatres

M. Nowak, P. Carbonez, M. Krauss,F. R. Verdun,J. Damet

SCIENTIFIC REPORTS(2020)

引用 6|浏览17
暂无评分
摘要
We used the Timepix3 hybrid pixel detector technology in order to determine the exposure of medical personnel to ionizing radiation in an interventional radiology room. We measured the energy spectra of the scattered radiation generated by the patient during X-ray image-guided interventional procedures. We performed measurements at different positions and heights within the theatre. We first observed a difference in fluence for each staff member. As expected, we found that the person closest to the X-ray tube is the most exposed while the least exposed staff member is positioned at the patient’s feet. Additionally, we observed a shift in energy from head to toe for practitioners, clearly indicating a non-homogenous energy exposure. The photon counting Timepix3 detector provides a new tool for radiation field characterisation that is easier-to-use and more compact than conventional X-ray spectrometers. The spectral information is particularly valuable for optimising the use of radiation protection gear and improving dosimetry surveillance programs. We also found the device very useful for training purposes to provide awareness and understanding about radiation protection principles among interventional radiology staff.
更多
查看译文
关键词
Applied physics,Health occupations,Techniques and instrumentation,Science,Humanities and Social Sciences,multidisciplinary