National Program for External Quality Assessment of Chinese Newborn Screening Laboratories.

Yuxuan Du,Wei Wang, Jiali Liu,Zhixin Zhang, Zhen Zhao,Falin He, Shuai Yuan,Zhiguo Wang

International journal of neonatal screening(2020)

引用 2|浏览7
暂无评分
摘要
OBJECTIVES:To analyze the coefficient of variation (CV) of external quality assessment (EQA) in Chinese newborn screening (NBS) laboratories. METHOD:EQA's robust CV was analyzed by the Clinet-EQA evaluation system. RESULTS:Participating laboratories of the EQA program increased annually. There was more than a 11-fold increase in phenylalanine (Phe) and thyroid stimulating hormone (TSH). It has shown a declining robust CV, which has tended to level off in recent years. The interquartile range (IQR) of Phe and TSH's robust CV has decreased from 15.5% to 1.5% and from 22.8% to 1.8%, respectively. Compared to bacterial inhibition assay (BIA), the robust CV of Phe has been shown to be relatively reduced in the fluorescence assay and quantitative enzymatic assay (QEA). The robust CV by ELISA was relatively unstable compared to DELFIA and FEIA. In addition, the robust CVs of glucose-6-phosphate dehydrogenase (G6PD) and 17-alpha-hydroxy progesterone (17-OHP) by Genetic Screening Processor (GSP) were lower than other systems. The median of robust CV by non-derivatized MS/MS (Fenghua) in Phe and free carnitine were around 2.2-4.7% and 2.6-5.2%. CONCLUSION:Neonatal screening has developed rapidly in China and the majority of participant laboratories had satisfactory performance for the quantitative results.