One-Step Implantation Of A 3d Neural Microelectrode Array

42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20(2020)

引用 1|浏览29
暂无评分
摘要
This paper presents a neurosurgical device called NEIT 2 (Nerve Electrode Insertion Tool) to implant a 3D microelectrode array into a peripheral nervous system. Using an elastomer-made nerve holder, the device is able to stable target a flexible nerve, and then safely inserts an electrode array into the fixed nerve. Finally, a nerve containment assembly is made at once. We conducted animal experiments to evaluate the proposed scenario using a 3D printed prototype and commercial microelectrodes. The results show that microelectrodes are successfully implanted into sciatic nerves of rats and neural signals are recorded through the chronically implanted electrodes.
更多
查看译文
关键词
Animals,Electrodes, Implanted,Microelectrodes,Rats,Sciatic Nerve