Impedance Plethysmography as an Alternative Measure of Reactive Hyperemia

42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20(2020)

引用 2|浏览6
暂无评分
摘要
Peripheral vascular flow in response to induced reactive hyperemia of the radial artery is used as a benchmark for non-invasive assessment of the endothelial function. As an alternative to standard modalities, this study investigates the suitability of impedance plethysmography to estimate peripheral vascular flow variations associated with the reactive hyperemia process. Results indicate a consistent variation of bio-impedance during the reactive hyperemia process at higher measurement frequencies and these variations are compatible with a standard tissue impedance model. Further, calculated features of bioimpedance has shown the capability of differentiating healthy and diabetic groups which is useful in estimating the endothelial dysfunction.
更多
查看译文
关键词
Diabetes Mellitus,Humans,Hyperemia,Plethysmography, Impedance,Radial Artery