Conventional pulse transit times as markers of blood pressure changes in humans

SCIENTIFIC REPORTS(2020)

引用 47|浏览28
暂无评分
摘要
Pulse transit time (PTT) represents a potential approach for cuff-less blood pressure (BP) monitoring. Conventionally, PTT is determined by (1) measuring (a) ECG and ear, finger, or toe PPG waveforms or (b) two of these PPG waveforms and (2) detecting the time delay between the waveforms. The conventional PTTs (cPTTs) were compared in terms of correlation with BP in humans. Thirty-two volunteers [50% female; 52 (17) (mean (SD)) years; 25% hypertensive] were studied. The four waveforms and manual cuff BP were recorded before and after slow breathing, mental arithmetic, cold pressor, and sublingual nitroglycerin. Six cPTTs were detected as the time delays between the ECG R-wave and ear PPG foot, R-wave and finger PPG foot [finger pulse arrival time (PAT)], R-wave and toe PPG foot (toe PAT), ear and finger PPG feet, ear and toe PPG feet, and finger and toe PPG feet. These time delays were also detected via PPG peaks. The best correlation by a substantial extent was between toe PAT via the PPG foot and systolic BP [− 0.63 ± 0.05 (mean ± SE); p < 0.001 via one-way ANOVA]. Toe PAT is superior to other cPTTs including the popular finger PAT as a marker of changes in BP and systolic BP in particular.