Acoustic Coupling Quantification in Ultrasound-Guided Focused Ultrasound Surgery: Simulation-Based Evaluation and Experimental Feasibility Study.

ULTRASOUND IN MEDICINE AND BIOLOGY(2020)

引用 3|浏览19
暂无评分
摘要
Adequate acoustic coupling between the therapeutic transducer and the patient's body is essential for safe and efficient focused ultrasound surgery (FUS). There is currently no quantitative method for acoustic coupling verification in ultrasound-guided FUS. In this work, a quantitative method was developed and a related metric was introduced: the acoustic coupling coefficient. This metric associates the adequacy of the acoustic coupling with the reflected signals recorded through an imaging probe during a low-energy sonication. The acoustic coupling issue was simulated in silico and validated through in vitro tests. Our results indicated a sigmoidal behavior of the introduced metric as the contact surface between the coupling system and the patient's skin increases. The proposed method could be a safety-check criterion for verifying the adequacy of the acoustic coupling before starting the FUS treatment, thus ensuring efficient energy transmission to the target and preventing damage to both the patient and the instrumentation. (E-mail: laura.morchi@santannapisa.it) (C) 2020 World Federation for Ultrasound in Medicine & Biology. All rights reserved.
更多
查看译文
关键词
Focused ultrasound surgery,High-intensity focused ultrasound treatment,Acoustic coupling verification,Ultrasound raw radiofrequency data,Patient safety