The Munich Valsalva Implantation Technique (Muvit) For Cardiac Output Reduction During Tevar: Vena Cava Occlusion With The Valsalva Maneuver

JOURNAL OF ENDOVASCULAR THERAPY(2021)

引用 19|浏览3
暂无评分
摘要
Purpose To demonstrate a physiologically induced alternative to the typical methods of reducing cardiac output during deployment of stent-grafts in the aortic arch and proximal aorta. Technique A modified Valsalva maneuver, the Munich Valsalva implantation technique (MuVIT), to raise the intrathoracic pressure, minimize backflow, and reduce the cardiac output is illustrated in a patient undergoing a triple-branch thoracic endovascular aortic repair (TEVAR). During manual mechanical ventilation, the adjustable pressure-limiting valve is carefully closed to 25 mm Hg, creating "manual bloating" of the lungs and sustained apnea. The increased intrathoracic pressure causes compression of the vena cava and pulmonary veins, reducing the venous backflow and gradually decreasing the arterial pressure. Once the desired pressure is obtained, the stent-graft is accurately deployed. The airway pressure is thereupon slowly reduced, and the patient is taken back to normal ventilation. The procedure is then finished following standard practice. Conclusion The MuVIT is a simple, noninvasive technique for cardiac output reduction during aortic arch TEVAR, eliminating the need for other invasive techniques.
更多
查看译文
关键词
aortic arch, cardiac output, deployment techniques, endograft, positive end-expiratory pressure, stent-graft, thoracic endovascular aortic repair, vena cava occlusion