An experimental evaluation of a concept to improve conventional aortic prostheses

Journal of Biomechanics(2020)

引用 2|浏览4
暂无评分
摘要
Conventionally used textile prosthesis for traditional open surgical repair (OSR) of aortic aneurysms have a lower compliance than the native aortic tissue. Graft placements lead to an acute drop in compliance which effects cardiovascular risk and the development of graft related complications. A custom-made spring casing was applied to a Dacron graft segment under physiological pressure conditions within a five-element biventricular mock circulation loop, to investigate experimentally a concept to improve the compliance of a conventional aortic prosthesis by changing the transverse graft cross-section. Two different prosthesis locations, proximal and distal of compliant silicone tubing were used to study uniaxial graft compression with an elastic device. To characterise the devices’ performance by means of pulse pressure (PP), diastolic pressure (Pdia) and pulse wave velocity(PWV), fluid pressures and flow were recorded.
更多
查看译文
关键词
Open surgical repair,Textile prosthesis,Dacron graft,Compliance,Pulse wave velocity,Mock circulation loop