Application of Molecularly Imprinted Polymer-modified Potentiometric Sensor for Quantitative Determination of Histamine in Serum

Analytical Sciences(2020)

引用 2|浏览4
暂无评分
摘要
A molecularly imprinted polymer-modified potentiometric histamine (HIS) sensor was prepared and used for quantitative determination of HIS in bovine serum. The calibration curve using the potential responses measured in 1 × 10 –3 mol L –1 phosphate buffer (pH 7.4) showed good linearity in the HIS concentration range of 3 × 10 –4 to 1 × 10 –2 mol L –1 ( r = 0.92), with a detection limit of 1.6 × 10 –4 mol L –1 . In bovine serum samples, the HIS sensor showed good recovery values of 91–104%. Therefore, this HIS sensor successfully determined the HIS concentration in bovine serum samples.
更多
查看译文
关键词
Keywords Potentiometric sensor, molecularly imprinted polymer, histamine, serum, quantitative determination, selectivity