A broadband multi distance approach to measure tissue oxygen saturation with continuous wave near infrared spectroscopy

European Conference on Biomedical Optics(2019)

引用 3|浏览17
暂无评分
摘要
Brain tissue oxygen saturation, StO 2, measured with near-infrared spectroscopy (NIRS) is of great clinical interest as it quantifies the balance between cerebral oxygen supply and demand. Some brain oximeters are based on spatially resolved spectroscopy (SRS), where NIRS data is collected at multiple distances from the light source to estimate a slope of light attenuation against distance. Other use a broadband approach which utilizes derivatives of the absorption spectra to estimate StO 2, such as broadband fitting (BF). We describe a novel algorithm, broadband spatially resolved spectroscopy (BB-SRS), for estimating StO 2. It is based on comparing the measured slope to a model of the attenuation slope, which depends on the optical properties of tissue. Fitting this model with a least squares fitting procedure recovers parameters describing absorption and scattering; the concentrations of oxy-and deoxy …
更多
查看译文
关键词
brain oximetry, near-infrared spectroscopy, broadband