Adverse Events Related to Cartiva Hemiarthroplasty of First Metatarsal: An Analysis of Reports to the United States Food and Drug Administration

FOOT & ANKLE SPECIALIST(2022)

引用 6|浏览7
暂无评分
摘要
Background: The results supporting Cartiva, a synthetic cartilage implant (Wright Medical) in hallux rigidus have come from limited institutions creating observational bias. Complications experienced in community centers are not routinely included in the published literature. To look at a broader range of potential complications, we reviewed the United States Food and Drug Administration's (FDA) voluntary device database and compared that data with published literature. Methods: The Manufacturer and User Facility Device Experience (MAUDE) database of the FDA was retrospectively reviewed between July 2016 and October 2019 using the product code: PNW, assigned for Cartiva. Results: A total of 49 events have been reported and implant subsidence was the most common with 16 reports. Others include fragmentation (9), infection (4), bone erosion (3), foreign body reaction (1) and unspecified (16). Thirty-five events mentioned further surgeries at a mean interval of 4.75 months. Conclusions: The analysis of the MAUDE database disclosed certain device-related dysfunctions that have been underreported in the published literature. Because of the voluntary nature of reporting, the true incidence of each complication is unknown with this data representing a baseline. The MAUDE database could be further strengthened by a more robust reporting mechanism or mandatory reporting of device-related complications.
更多
查看译文
关键词
hallux rigidus, FDA, Cartiva, synthetic cartilage implant, hydrogel, complications, MAUDE database
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要