Sabrina Buoro_Layout 1

Michela Seghezzi,Sabrina Buoro,Giulia Previtali, Valentina Moioli, Barbara Manenti, Ramon Simon-Lopez,Cosimo Ottomano,Giuseppe Lippi

semanticscholar(2018)

引用 0|浏览6
暂无评分
摘要
Background: The cell population data (CPD) measured by Sysmex XN-9000 can be used for screening many hematological and non-hematological disorders. Since little in formation is available on harmonization of CPD among different instrumentation and clinical laboratories, this study aimed at assessing the current degree of CPD harmonization between separate Sysmex XN modules allocated to the same laboratory. Methods: A total number of 78291 data were used for verification of within-run imprecision, analyzers harmoni zation, reference ranges and assessment of blood sample stability of CPD parameters, including results of daily quality control testing and those generated in samples collected from blood donors and healthy volunteers. Results: Within-run imprecision of CPD parameters ranged between 0.4 and 14.1%. Good agreement was found among five different XN-modules, especially when values were adjusted after calculation of instrument-specific alignment factors. The bias of all parameters remained always Kratak sadr`aj