Comparison of Clinical Performance of VectorFlow and Palindrome Symmetric-Tip Dialysis Catheters: A Multicenter, Randomized Trial.

Journal of Vascular and Interventional Radiology(2020)

引用 2|浏览17
暂无评分
摘要
The 90-day primary patency rates of Palindrome and VectorFlow catheters were not significantly different, and both achieved sustained high Q through 90 day follow-up. However, dialysis adequacy based on Kt/V was consistently better with the VectorFlow catheter versus the Palindrome.
更多
查看译文
关键词
BMI,ESRD,Kt/V,QB,URR
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要