Comparison of Clinical Performance of VectorFlow and Palindrome Symmetric-Tip Dialysis Catheters: A Multicenter, Randomized Trial.
Journal of Vascular and Interventional Radiology(2020)
摘要
The 90-day primary patency rates of Palindrome and VectorFlow catheters were not significantly different, and both achieved sustained high Q through 90 day follow-up. However, dialysis adequacy based on Kt/V was consistently better with the VectorFlow catheter versus the Palindrome.
更多查看译文
关键词
BMI,ESRD,Kt/V,QB,URR