Fluispotter, A Novel Automated And Wearable Device For Accurate Volume Serial Dried Blood Spot Sampling
BIOANALYSIS(2020)
摘要
Aim: A novel automated serial dried blood spot (DBS) sampler, 'Fluispotter', was tested for its sampling performance. Materials & methods: An LC-MS/MS method was developed for the analysis of cortisol in DBS samples serially spotted by Fluispotter. The cortisol concentrations in 148 paired DBS and plasma samples were compared across a hematocrit (HCT) range of 22-55%. Results: The interassay accuracy and precision were <10%. Overall assay bias was negligible across the HCTs tested when analyzing the whole-spot DBS samples. The accuracy and precision of the blood volume in 10 mu l DBS samples spotted by Fluispotters and micropipettes were within 3%. Deming regression and Bland-Altman analysis showed a good agreement of DBS-predicted and measured plasma cortisol. Conclusion: The Fluispotter performed serial sampling with high accuracy and precision of the sample blood volume.
更多查看译文
关键词
cortisol, dried blood spot, fluispotter, hematocrit, LC-MS/MS, microsampling
AI 理解论文