A Pilot Study Comparing Aortic Valve Area Estimates Derived from Fick Cardiac Output with Estimates Based on Cheetah-NICOM Cardiac Output

SCIENTIFIC REPORTS(2020)

引用 3|浏览7
暂无评分
摘要
Cardiac output during cardiac catheterization is often estimated using the modified Fick method (CO Fick ). In this proof-of-concept, prospective non-randomized study carried out in a single academic healthcare centre, we examined whether replacing CO Fick in the Gorlin formula with Cheetah-NICOM monitor cardiac output (CO Cheetah ) could produce an accurate and precise estimate of aortic valve area in patients with severe aortic stenosis. In twenty-six subjects, CO Fick and CO Cheetah were obtained concurrently. A spot and 3-minute running average of CO Cheetah was used. Bland and Altman analysis was used to derive bias, 95% limits of agreement (LOA) and confidence intervals (CI). The mean difference (bias) between AVA Cheetah (average) and AVA Fick was 0.11 cm 2 and the 95% LOA were ±0.42 cm 2 . The 95% CI of the bias was 0.02–0.2 cm 2 . The bias and 95% LOA of AVA Cheetah (spot value) were 0.14 ± 0.42cm 2 , with a 95% CI of 0.06–0.23 cm 2 . No proportional bias was present. AVA Cheetah thus appears to be a reasonably accurate measure of AVA in patients with severe aortic stenosis compared to AVA Fick measured using a modified Fick CO. However, the limits of agreement were not narrow enough to consider AVA Cheetah and AVA Fick interchangeable.
更多
查看译文
关键词
Cardiology,Interventional cardiology,Science,Humanities and Social Sciences,multidisciplinary
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要