A cracking complication: Displacement of existing clavicle fracture complicating subcutaneous implantable cardioverter defibrillator implantation.

Indian Pacing and Electrophysiology Journal(2020)

引用 0|浏览14
暂无评分
摘要
An 18 year old male with an incompletely healed clavicle fracture presented with unexplained syncope. Subsequent investigations were consistent with a diagnosis of arrhythmogenic right ventricular cardiomyopathy (ARVC). A subcutaneous implantable cardioverter-defibrillator (S-ICD) was successfully implanted and defibrillation threshold (DFT) testing performed as per standard protocol. Shortly following the procedure, the patient complained of pain and swelling over the left clavicle. A radiograph revealed aggravation and displacement of the underlying clavicle fracture. Surgical reduction and internal fixation was performed one week later.
更多
查看译文
关键词
Subcutaneous implantable cardioverter defibrillator,Clavicle,Shoulder injuries,Defibrillation threshold testing