Establishment of national diagnostic reference levels in dental cone beam computed tomography in Switzerland.

Marine Deleu,Dorothea Dagassan,Isabelle Berg, Julie Bize, Karl Dula,Vincent Lenoir, Barbara Ott,Julien G Ott,Francis R Verdun, Robert Weber, Panagiotis Ziglinas, Cesare Zingariello,Marta Sans Merce

Dento maxillo facial radiology(2020)

引用 13|浏览9
暂无评分
摘要
OBJECTIVES:The aim of this study was to establish diagnostic reference levels (DRLs) in the field of dental maxillofacial and ear-nose-throat (ENT) practices using cone beam CT (CBCT) in Switzerland. METHODS:A questionnaire was sent to owners of CBCTs in Switzerland; to a total of 612 institutions. The answers were analyzed for each indication, provided that enough data were available. The DRLs were defined as the 75th percentile of air kerma product distribution (PKA). RESULTS:227 answers were collected (38% of all centers). Third quartile of PKA values were obtained for five dental indications: 662 mGy cm² for wisdom tooth, 683 mGy cm² for single tooth implant treatment, 542 mGy cm² for tooth position anomalies, 569 mGy cm² for pathological dentoalveolar modifications, and 639 mGy cm² for endodontics. The standard field of view (FOV) size of 5 cm in diameter x 5 cm in height was proposed. CONCLUSIONS:Large ranges of FOV and PKA were found for a given indication, demonstrating the importance of establishing DRLs as well as FOV recommendations in view of optimizing the present practice. For now, only DRLs for dental and maxillofacial could be defined; because of a lack of ENT data, no DRL values for ENT practices could be derived from this survey.