An Automated SeaFAST ICP-DRC-MS Method for the Determination of 90 Sr in Spent Nuclear Fuel Leachates.

MOLECULES(2020)

引用 5|浏览5
暂无评分
摘要
To reduce uncertainties in determining the source term and evolving condition of spent nuclear fuel is fundamental to the safety assessment. ss -emitting nuclides pose a challenging task for reliable, quantitative determination because both radiometric and mass spectrometric methodologies require prior chemical purification for the removal of interfering activity and isobars, respectively. A method for the determination of Sr-90 at trace levels in nuclear spent fuel leachate samples without sophisticated and time-consuming procedures has been established. The analytical approach uses a commercially available automated pre-concentration device (SeaFAST) coupled to an ICP-DRC-MS. The method shows good performances with regard to reproducibility, precision, and LOD reducing the total time of analysis for each sample to 12.5 min. The comparison between the developed method and the classical radiochemical method shows a good agreement when taking into account the associated uncertainties.