Real-Time Thermoacoustic Imaging And Thermometry During Focused Microwave Heating In Multilayer Breast Phantom

2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS)(2019)

引用 4|浏览8
暂无评分
摘要
This study describes real-time thermoacoustic imaging (TAI) and thermometry (TAT) for feedback during thermal therapies. The system uses a linear ultrasound array and imaging system to noninvasively estimate temperature. The system was validated in gel phantoms, muscle and fat tissue samples. A mean error between 10 and 20% per degree was observed for estimating temperature. A TAT system using a single element ultrasound probe was also integrated with a focused microwave therapy (FMT) setup for generating local heating, leading to a mean error of 1.18 +/- 0.87 degrees C compared to thermocouples placed near the region of interest. The results support the long-term goal of developing a closed-loop platform for image-guided ablation therapy for breast cancer.
更多
查看译文
关键词
radiofrequency ablation, hyperthermia, focused microwave therapy, breast cancer