Wafer Level Characterization Of Row-Column Addressed Cmut Arrays

2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS)(2019)

引用 1|浏览24
暂无评分
摘要
This paper presents a measurement methodology for wafer level characterization of row-column addressed (RCA) capacitative micromachined ultrasound transducers (CMUT). Characterization of a 62+62 element RCA CMUT is presented. To facilitate wafer level electrical characterization measurements between adjacent electrodes can be used to characterize the device. This allows for determination of the individual element capacitance. Current-voltage measurements between adjacent top or bottom electrodes provides valuable information about process yield.
更多
查看译文
关键词
measurement methodology,row-column addressed capacitative micromachined ultrasound transducers,62+62 element RCA CMUT,wafer level electrical characterization measurements,element capacitance,current-voltage measurements,row-column addressed CMUT arrays,adjacent electrodes
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要