Wafer Level Characterization Of Row-Column Addressed Cmut Arrays
2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS)(2019)
摘要
This paper presents a measurement methodology for wafer level characterization of row-column addressed (RCA) capacitative micromachined ultrasound transducers (CMUT). Characterization of a 62+62 element RCA CMUT is presented. To facilitate wafer level electrical characterization measurements between adjacent electrodes can be used to characterize the device. This allows for determination of the individual element capacitance. Current-voltage measurements between adjacent top or bottom electrodes provides valuable information about process yield.
更多查看译文
关键词
measurement methodology,row-column addressed capacitative micromachined ultrasound transducers,62+62 element RCA CMUT,wafer level electrical characterization measurements,element capacitance,current-voltage measurements,row-column addressed CMUT arrays,adjacent electrodes