DEVELOPMENT OF A NEW SECONDARY STANDARD FOR H*(10).

K Bairlein,B Behnke,O Hupe

RADIATION PROTECTION DOSIMETRY(2020)

引用 0|浏览6
暂无评分
摘要
A secondary standard for ambient dose equivalent, H-*(10), is necessary for the dissemination of the unit Sievert (Sv), but there is no such standard commercially available currently. Furthermore, the measurement of H-*(10) instead of calculating H-*(10) from air kerma and conversion coefficients is needed for unknown radiation fields. We developed a prototype of a new secondary standard for H-*(10) based on a spherical 1 l ionization chamber for air kerma. This chamber was modified with copper wires at the inner surface to adjust the response of the chamber according to H-*(10). Additionally, a Makrolon shell and an aluminium coating were added to optimize the response at energies below 50 keV. The prototype fulfils the requirements given in ISO 4037-2 in the energy range from 12 keV to 7 MeV. In combination with an electrometer, it can be used as area dosemeter, suitable for pulsed fields and for low energy radiation.