Clinical Magnetic Resonance Imaging of Arthroplasty at 1.5T.

JOURNAL OF ORTHOPAEDIC RESEARCH(2020)

引用 8|浏览22
暂无评分
摘要
Magnetic resonance imaging (MRI) has historically been avoided for the routine clinical evaluation of metal implants at many clinical centers due to the presence of artifact that creates in-plane and through-plane distortions and signal intensity voids in generated images. However, when the image acquisition parameters are appropriately modified and advanced multi-spectral pulse sequences are used, high-quality diagnostic images can be generated and may be used for diagnosing patients with suspected periprosthetic pathology. MRI provides superior soft-tissue contrast and excellent sensitivity for mobile water and is, therefore, a valuable tool in the evaluation of these patients, given the increasing prevalence of arthroplasty within the general population. Knowledge of expected normal postoperative appearance in patients with total hip arthroplasty, total knee arthroplasty, and total shoulder arthroplasty facilitates the detection of abnormal findings in this population, as does familiarity with common pathologic conditions encountered in the periprosthetic region. This review article will provide background information regarding the presence of image artifacts, methods to reduce the artifacts, and application of MRI at 1.5 T for evaluating common complications in subjects with total knee arthroplasty, total hip arthroplasty, and total shoulder arthroplasty.
更多
查看译文
关键词
adverse soft tissue reaction,arthroplasty,arthroplasty-hip,arthroplasty-knee,diagnostic imaging,implant fixation,implant wear