Clinical method evaluation of hemoglobin S and C identification by top-down selected reaction monitoring and electron transfer dissociation

Clinical Proteomics(2019)

引用 3|浏览43
暂无评分
摘要
Background Biological diagnosis of hemoglobin disorders is a complex process relying on the combination of several analytical techniques to identify Hb variants in a particular sample. Currently, hematology laboratories usually use high-performance liquid chromatography (HPLC), capillary electrophoresis and gel-based methods to characterize Hb variants. Co-elution and co-migration may represent major issues for precise identification of Hb variants, even for the most common ones such as Hb S and C. Methods We adapted a top-down selected reaction monitoring (SRM) electron transfer dissociation (ETD) mass spectrometry (MS) method to fit with a clinical laboratory environment. An automated analytical process with semi-automated data analysis compatible with a clinical practice was developed. A comparative study between a reference HPLC method and the MS assay was performed on 152 patient samples. Results The developed workflow allowed to identify with high specificity and selectivity the most common Hb variants (Hb S and Hb C). Concordance of the MS-based approach with HPLC was 71/71 (100%) for Hb S and 11/11 (100%) for Hb C. Conclusions This top-down SRM ETD method can be used in a clinical environment to detect Hb S and Hb C.