Electron radiotherapy (IOERT) for applications outside of the breast: Dosimetry and influence of tissue inhomogeneities.

Physica Medica(2020)

引用 5|浏览61
暂无评分
摘要
•In-phantom commissioning of the Mobetron was performed and externally validated.•We measured dose enhancement factor of bone in IOERT up to 6% with energy 12 MeV and 3 cm applicator.•In vivo dosimetry on 126 patients for treatments out of breast resulted in average +3.1% dose difference.•Discrepancies of in vivo to prescribed doses may be also explained with presence of inhomogeneous tissues.
更多
查看译文
关键词
Intra-operative,Radiation therapy,Dosimetry,Radiochromic film,Electron therapy