Central Aortic Blood Pressure Estimation In Children And Adolescents: Results Of The Kidcorebp Study

JOURNAL OF HYPERTENSION(2020)

引用 29|浏览18
暂无评分
摘要
Background: Central aortic SBP (cSBP) may have superior prognostic value compared with peripheral SBP (pSBP), but noninvasive cSBP measurement techniques have not been formally validated in children and adolescents.Method: This study assessed the accuracy of two automated devices and the radial tonometry/transfer function method (RT-TF) for estimating central pressures and pulse pressure amplification (PPA) in this population, with adherence to validation guidelines for central pressure devices. In 69 children/adolescents aged 3-18 years undergoing clinically indicated aortic catheterization, high fidelity ascending aortic cSBP was measured with a micromanometer-tipped wire and compared with values from SphygmoCor XCEL, Mobil-O-Graph (systolic/diastolic calibration, MoG-C1, or mean/diastolic calibration, MoG-C2) and RT-TF. Reference intra-arterial pSBP was derived from the tonometry pulse calibrated to central mean/diastolic pressures.Results: XCEL, MoG-C1 and MoG-C2 overestimated cSBP by 7.9 +/- 6.8 mmHg (mean +/- SD), 5.7 +/- 10.3 mmHg, and 19.1 +/- 14.9 mmHg, exceeding the validation cut-off (5 +/- 8 mmHg). Brachial pSBP was also overestimated by XCEL (10.9 +/- 8.4 mmHg) and Mobil-O-Graph (11.5 +/- 12.3 mmHg). By contrast, central and brachial diastolic pressures were underestimated by the automated devices, albeit mostly within acceptable limits; pulse pressures were, therefore, substantially overestimated. Central-brachial PPA (4.5 +/- 4.4 mmHg) was overestimated by XCEL (8.7 +/- 3.2 mmHg) and MoG-C1 (11.1 +/- 6.4 mmHg), but underestimated by MoG-C2 (-3.0 +/- 6.6 mmHg). Given accurate pulse calibration, RT-TF achieved acceptable accuracy for cSBP (-0.2 +/- 4.6 mmHg) and central-radial PPA (1.9 +/- 5.1 mmHg).Conclusion: In conclusion, XCEL and Mobil-O-Graph overestimated pSBP and cSBP in children and adolescents. cSBP can be obtained via the same transfer function used in adults, but accurate pressure pulse calibration is critical.
更多
查看译文
关键词
aorta, central blood pressure, children, device, hypertension, validation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要