Efficiency enhanced electrolysis-based tritium continuous monitor
Journal of Radioanalytical and Nuclear Chemistry(2019)
摘要
Developing an online monitoring system becomes useful for continuously detection of tritium in water. A plastic scintillator chamber was used to tritium detection in hydrogen gas produced by a proton exchange membrane electrolysis cell. The fractionation factor (proton selectivity) was estimated as being of 5.41 ± 0.20, and detection efficiency of the plastic scintillator chamber was 4.25 ± 0.22%. For liquid samples, the minimum detectable activity of the developed system was estimated as 655 kBq L −1 . Using this online technique for tritium monitoring based on electrolysis is expected to be useful for analysis of large tritiated water bearing samples.
更多查看译文
关键词
Proton exchange membrane, Electrolysis, Tritium, Continuous monitoring, Fractionation factor