Flat-field calibration method for hyperspectral frame cameras

METROLOGIA(2019)

引用 7|浏览8
暂无评分
摘要
This paper presents a method for characterising spatial responsivity of hyperspectral cameras. Knowing the responsivity of the camera as a function of pixel coordinates allows applying a flat-field correction on image data. The method is based on scanning the field of view of the camera with a broadband radiance source, based on an integrating sphere, and combining the captured frames to synthesise a uniform radiance source filling the whole field of view of the camera at the focus distance. The method was compared with a traditional approach where the aperture of an integrating sphere is imaged from a close distance, filling the entire field of view of the camera. The measurement setup was tested with a hyperspectral camera, based on a tunable Fabry-Perot interferometer. Without the flat-field correction, the average standard deviation of the pixel responsivities across all the spectral channels of the camera was 3.78%. After the correction, the average standard deviation was reduced to 0.40% and 3.25% for the aperture-scanning method and the close-distance method, respectively. The expanded uncertainty (k = 2) for the flat-field correction obtained using the scanning method was 0.68%-0.78%, depending on the spectral channel of the camera.
更多
查看译文
关键词
hyperspectral,camera,calibration,flat-fielding,vignetting,Fabry-Perot,measurement uncertainty
AI 理解论文