Denoising of diffusion MRI improves peripheral nerve conspicuity and reproducibility.

JOURNAL OF MAGNETIC RESONANCE IMAGING(2020)

引用 7|浏览46
暂无评分
摘要
Background Quantitative diffusion MRI is a promising technique for evaluating peripheral nerve integrity but low signal-to-noise ratio (SNR) can impede measurement accuracy. Purpose To evaluate principal component analysis (PCA) and generalized spherical deconvolution (genSD) denoising techniques to improve within-subject reproducibility and peripheral nerve conspicuity. Study Type Prospective. Subjects Seven healthy volunteers and three peripheral neuropathy patients. Field Strength/Sequence 3T/multiband single-shot echo planar diffusion sequence using multishell 55-direction scheme. Assessment Images were processed using four methods: "original" (no denoising), "average" (10 repetitions), "PCA-only," and "PCA + genSD." Tibial and common peroneal nerve segmentations and masks were generated from volunteer diffusion data. Quantitative (SNR and contrast-to-noise ratio [CNR]) values were calculated. Three radiologists qualitatively evaluated nerve conspicuity for each method. The two denoising methods were also performed in three patients with peripheral neuropathies. Statistical Tests For healthy volunteers, calculations included SNR and CNRFA (computed using FA values). Coefficient of variation (CV%) of CNRFA quantified within-subject reproducibility. Groups were compared with two-sample t-tests (significance P < 0.05; two-tailed, Bonferroni-corrected). Odds ratios (ORs) quantified the relative rates of each of three radiologists confidently identifying a nerve, per slice, for the four methods. Results "PCA + genSD" yielded the highest SNR (mean(overall) = 14.83 +/- 1.99) and tibial and common peroneal nerve CNRFA (mean(tibial) = 3.45, mean(peroneal) = 2.34) compared to "original" (P- SNR < 0.001; P- CNR = 0.011) and "PCA-only" (P- SNR < 0.001, P- CNR < 0.001). "PCA + genSD" had higher within-subject reproducibility (low CV%) for tibial (6.04 +/- 1.98) and common peroneal nerves (8.27 +/- 2.75) compared to "original" and "PCA-only." The mean FA was higher for "original" than "average" (P < 0.001), but did not differ significantly between "average" and "PCA + genSD" (P = 0.14). "PCA + genSD" had higher tibial and common peroneal nerve conspicuity than "PCA-only" (ORtibial = 2.50, P < 0.001; ORperoneal = 1.86, P < 0.001) and "original" (ORtibial = 2.73, P < 0.001; ORperoneal = 2.43, P < 0.001). Data Conclusion PCA + genSD denoising method improved SNR, CNRFA, and within-subject reproducibility (CV%) without biasing FA and nerve conspicuity. This technique holds promise for facilitating more reliable, unbiased diffusion measurements of peripheral nerves. Technical Efficacy Stage: 1 J. Magn. Reson. Imaging 2020;51:1128-1137.
更多
查看译文
关键词
diffusion tensor imaging,peripheral nerve,denoising,MRI,3T
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要