This report documents measurements of cosmic ray induced background neutrons using a new Bonner sphere spectrometer (BSS) based on a 3He thermal neutron counter (SP9 type). Three spheres with a lead or copper layer were added to a BSS to enhance its sensitivity in the high-energy range above 20 MeV and enable the spectrometer to cover the entire energy range from thermal to multi-GeV. The response function of the spectrometer was calculated using the Monte Carlo code MCNP6. Below 20 MeV, the responses of polyethylene spheres were experimentally verified with 6 mono-energetic neutron sources. A neutron spectrum was measured near the ground at one location in Beijing. Also deduced from the spectrum were the rates of neutron fluence, ambient dose equivalent, and effective dose.
更多
查看译文
关键词
Cosmic ray,Bonner sphere spectrometer,Response,MCNP6,Neutron spectrum