Diagnostic Performance Using Automated Breast Ultrasound System for Breast Cancer in Chinese Women Aged 40 Years or Older: A Comparative Study.

Ultrasound in Medicine & Biology(2019)

引用 12|浏览14
暂无评分
摘要
The purpose of this study was to investigate the diagnostic performance of the automated breast ultrasound system (ABUS) compared with hand-held ultrasonography (HHUS) and mammography (MG) for breast cancer in women aged 40 y or older. A total of 594 breasts in 385 patients were enrolled in the study. HHUS, ABUS and MG exams were performed for these patients. Follow-up and pathologic findings were used as the reference standard. Based on the reference standard, 519 units were benign or normal and 75 were malignant. The sensitivity, specificity, accuracy and Youden index were 97.33%, 89.79%, 90.74% and 0.87 for HHUS; 90.67%, 92.49%, 92.26% and 0.83 for ABUS; 84.00%, 92.87%, 91.75% and 0.77 for MG, respectively. The specificity of ABUS was significantly superior to that of HHUS (p = 0.024). The area under the receiver operating characteristic curve was 0.936 for HHUS, which was the highest, followed by 0.916 for ABUS and 0.884 for MG. However, the difference was not statistically significant (p > 0.05). In conclusion, the diagnostic performance of ABUS for breast cancer was equivalent to HHUS and MG and potentially can be used as an alternative method for breast cancer diagnosis.
更多
查看译文
关键词
Breast cancer,Hand-held ultrasonography,Automated breast ultrasound system,Mammography