Allergic contact dermatitis from glucose sensors in type 1 diabetes patients.

CONTACT DERMATITIS(2019)

引用 74|浏览0
暂无评分
摘要
Background Allergic contact dermatitis caused by glucose sensors has become an increasing problem. Contact allergies to isobornyl acrylate, colophonium, ethyl cyanoacrylate and N,N-dimethylacrylamide have been reported. However, there is a paucity of information regarding the prevalence of sensor-related dermatitis and detailed patient histories. Objectives To evaluate diabetes patients who have developed allergic contact dermatitis caused by glucose sensors. Patients and methods Seventy patients with suspected contact allergy to glucose sensors referred to university hospital dermatology clinics in southern Finland were patch tested with the baseline series, an isobornyl acrylate dilution series, and a number of other acrylates. Atopic constitution, contact allergies, the severity of the dermatitis and the ability to continue with the same or another type of sensor were recorded. Results Positive patch test reactions to isobornyl acrylate were seen in 51 of 63 (81%) Freestyle Libre users. Colophonium or its derivatives gave positive reactions in five of seven Medtronic Enlite users. The median duration of sensor use before dermatitis was 6 months. The estimated prevalences of contact allergy to sensor adhesives were 0.7% for Freestyle Libre and 0.8% for Enlite. Conclusions We suggest that patients who have difficulties in finding a tolerable glucose sensor or an insulin pump should be patch tested.
更多
查看译文
关键词
acrylates,adhesives,allergic contact dermatitis,CAS no,5888-33-5,glucose sensors,isobornyl acrylate,medical device,type 1 diabetes
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要