Energy Calibration Of High-Resolution X-Ray Tes Microcalorimeters With 3 Ev Optical Photons

IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY(2019)

引用 10|浏览73
暂无评分
摘要
With the improving energy resolution of transition-edge sensor (TES) based microcalorimeters, performance verification and calibration of these detectors have become increasingly challenging, especially in the energy range below 1 keV where fluorescent atomic X-ray lines have linewidths that are wider than the detector energy resolution and require impractically high statistics to determine the gain and deconvolve the instrumental profile. Better behaved calibration sources such as grating monochromators are too cumbersome for space missions and are difficult to use in the lab. As an alternative, we are exploring the use of pulses of 3 eV optical photons delivered by an optical fiber to generate combs of known energies with known arrival times. Here, we discuss initial results of this technique obtained with 2 and 0.7 eV resolution Xray microcalorimeters. With the 2 eV detector, we have achieved photon number resolution for pulses with mean photon number up to 133 (corresponding to 0.4 keV).
更多
查看译文
关键词
Calibration sources, Energy resolution, Microcalorimeters, Transition edge sensors, X-ray detectors