Characterisation of a small electrode HPGe detector

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2019)

引用 4|浏览32
暂无评分
摘要
Small electrode HPGe detectors in an inverted coaxial geometry are increasingly in use in applications where both high efficiency and excellent energy resolution are required. The unusual electric field configuration of these detectors results in extremely long charge collection times compared to planar and coaxial devices. In this work we have characterised such a detector using gamma-ray coincidence measurements and optimised an electric field simulation to reproduce the positional variation of detector response. We show that, alongside accurate crystal geometry and applied electric potential, a temperature correction is crucial to correctly determining appropriate charge carrier mobility parameters. This work will help to guide the future development of HPGE detectors for applications including radioactive waste assay, radio-isotope dating, and fundamental nuclear physics.
更多
查看译文
关键词
HPGe,SAGe Well,Charge collection,Charge carrier mobility