Mechanical Active Compression-Decompression With Lucas2ad Provide Equivocal P(Et)Co(2) Values To The Standard Lucas2 During Out-Of-Hospital Cardiopulmonary Resuscitation

Circulation(2017)

引用 23|浏览30
暂无评分
摘要
Introduction: Studies comparing manual active compression-decompression CPR (ACD-CPR) to manual CPR have shown similar survival rates. Recently a new mechanical ACD-CPR device (LUCAS2AD, Jolife AB/Physio-Control, Lund, Sweden) was compared to the standard LUCAS2 device in pigs, documenting improved hemodynamics. In this first clinical study on mechanical ACD-CPR we hypothesized that it would produce better hemodynamic variables than the standard LUCAS2 device (ClinicalTrials.gov NCT02479152). Methods: The trial is a prospective block-randomized unblinded study of out-of-hospital cardiac arrest patients u003e18 years old with anterior-posterior (AP) chest diameter 185-280mm, treated by the physician manned vehicle in Oslo. Inclusion criteria: simultaneous minimum 5 minutes of LUCAS2/LUCAS2AD chest compressions, intubated and side stream pCO2 measurements. Exclusion criteria: traumatic arrest, pregnancy, imprisonment and previous chest surgery. LIFEPAK 15 data was exported via CodeStat 10 (both PhysioControl, R...
更多
查看译文
关键词
Cardiac arrest, Cardiopulmonary resuscitation, Hemodynamics, Advanced life support, Resuscitation