Characterizing The Electron Response And Position Sensitivity For Radiation In Plastic Scintillators

2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC)(2017)

引用 0|浏览11
暂无评分
摘要
To develop a tissue-equivalent dosimeter based on LET measurement, we examine the characteristics of a candidate material, plastic scintillators. The energy resolutions of plastic scintillators are examined by using the measured electron responses defined as the relative light output per unit energy of three plastic scintillators EJ-200, EJ-212, and EJ-252. The position sensitivity characteristic is examined by using a square-aligned plastic scintillator rod, and multi-segmented photomultiplier tubes are attached to both ends of the respective rod. The relation between signals from both end PMTs and the position of radiation incidence in the rod are obtained.
更多
查看译文
关键词
measured electron responses,EJ-212,EJ-252,position sensitivity characteristic,square-aligned plastic scintillator rod,tissue-equivalent dosimeter,LET measurement,energy resolutions,multisegmented photomultiplier tubes,radiation incidence