Calibration Of Gain Ratios On Nv-Meters With The Reference Step Method
2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018)(2018)
摘要
The Reference Step Method for calibrators can be modified for calibration also of the gain ratios on a meter. The method has earlier been evaluated in the range 100 mV - 1000 V on DMM HP 3458A1 and shows that an accuracy of typically < 0.2 mu V/V could be obtained. We have investigated the performance of the method when calibrating ratios in the range 1 mV -100 mV on two common nV-meters. In our comparisons with the Josephson Voltage Standard the differences are within +/- 12 mu V/V at the ratio 10mV:1mV and within +/- 1.3 mu V/V at 100mV:10mV.
更多查看译文
关键词
DC voltage ratio, dc voltage scale, dc voltage build-up, low voltage measurements, nanovoltmeter