Diode‐based transmission detector for IMRT delivery monitoring: a validation study

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS(2016)

引用 22|浏览17
暂无评分
摘要
The purpose of this work was to evaluate the potential of a new transmission detector for real-time quality assurance of dynamic-MLC-based radiotherapy. The accuracy of detecting dose variation and static/dynamic MLC position deviations was measured, as well as the impact of the device on the radiation field (surface dose, transmission). Measured dose variations agreed with the known variations within 0.3%. The measurement of static and dynamic MLC position deviations matched the known deviations with high accuracy (0.7-1.2 mm). The absorption of the device was minimal (similar to 1%). The increased surface dose was small (1%-9%) but, when added to existing collimator scatter effects could become significant at large field sizes (>= 30 x 30 cm(2)). Overall the accuracy and speed of the device show good potential for real-time quality assurance.
更多
查看译文
关键词
transmission detector,real-time quality assurance,IMRT quality assurance,in vivo dosimetry