Low-Power, High Data-Rate Digital Capsule Endoscopy Using Human Body Communication

APPLIED SCIENCES-BASEL(2018)

引用 26|浏览17
暂无评分
摘要
A technology for low-power high data-rate digital capsule endoscopy with human body communication (CEHBC) is presented in this paper. To transfer the image data stably with low power consumption, the proposed system uses three major schemes: Frequency selective digital transmission (FSDT) modulation with HBC, the use of an algorithm to select electrode pairs, and the LineSync algorithm. The FSDT modulation supports high-data rate transmission and prevents the signal attenuation effect. The selection algorithm of the electrode pair finds the best receiving channel. The LineSync algorithm synchronizes the data and compensates for data polarity during the long data transmission section between the capsule endoscope and the receiver. Because all the major functional blocks of the CEHBC transmitter can be implemented as digital logics, they can be easily fabricated using the field programmable gate array (FPGA). Moreover, this CEHBC transmitter can achieve low power-consumption and can support a relatively high data rate in spite of using its clock a few tens of MHz slower. The proposed CEHBC-TXD is the digital portion of the CEHBC transmitter that provides low-power (3.7 mW) and high data-rate (6 Mbps) performance while it supports a high-resolution image (480 x 480 byte) at 3.13 fps.
更多
查看译文
关键词
capsule endoscopy,human body communication,frequency selective digital communication,selection algorithm of electrode pair,synchronization