Vibration Mitigation for a Cryogen-Free Dilution Refrigerator for the AMoRE-Pilot Experiment

Journal of Low Temperature Physics(2018)

引用 9|浏览28
暂无评分
摘要
The Advanced Mo-based Rare process Experiment utilizes a cryogen-free dilution refrigerator to operate its low-temperature detectors. Mechanical vibration originating from its pulse tube refrigerator can affect the detector performance. A mechanical filter system has been installed between the 4K and still plates with eddy current dampers in addition to a spring-loaded damping system previously installed below the mixing chamber plate of the cryostat. The filters significantly mitigated vibrations and improved the detector signals.
更多
查看译文
关键词
Calorimeters,Double beta decay detectors,Cryocoolers,Cryogenics,Dark matter detectors