Use of k Z -space for high through-plane resolution in multislice MRI: Application to prostate.

MAGNETIC RESONANCE IN MEDICINE(2019)

引用 4|浏览23
暂无评分
摘要
Purpose: The goal of this work is to demonstrate 1 mm through-plane resolution in multislice T2SE MRI using k(Z)-space processing of overlapping slices and show applicability in prostate MRI. Methods: Multiple overlapped slices are acquired and Fourier transformed in the slice-select direction. The slice profile is incorporated into a Tikhonov-regularized reconstruction. Through-plane resolution is tested in a resolution phantom. An anthropomorphic prostate phantom is used to study the SNR, and results are compared with theoretical prediction. The proposed method is tested in 16 patients indicated for clinical prostate MRI who gave written informed consent as overseen by our IRB. The "proposed" vs. "reference" multislice images are compared using multiple evaluation criteria for through-plane resolution. Results: The modulation transfer function (MTF) plots of the resolution phantom show good modulation at frequency 0.5 lp/mm, demonstrating 1 mm through-plane resolution restoration. The SNR measurements experimentally match the theoretically predicted values. The radiological evaluation shows that the proposed method is superior to the reference method for five criteria of sharpness but inferior with respect to artifacts. Conclusions: In conjunction with overlapped slices a k(Z)-space-based reconstruction approach can be used to improve through-plane resolution in multislice T2SE MRI. 1 mm resolution is demonstrated from 3.2 mm thick slices. The in vivo results from prostate MRI show improved sharpness when compared to the standard multislice method.
更多
查看译文
关键词
MRI,multislice,reconstruction,through plane resolution,T2-weighted Spin Echo
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要