Experimental characterization of a long counter for neutron fluence measurement

Radiation Measurements(2018)

引用 3|浏览85
暂无评分
摘要
A De Pangher-type long counter was constructed at Peking University for neutron fluence measurement. The responses and effective centers of the long counter were calculated using the Monte Carlo method. The variations with neutron energy of the position of the effective center, calculated by the Monte Carlo code, was experimentally validated with a 241AmBe neutron source and several mono-energetic neutron sources ranging from 100 keV to 6 MeV. Long counter calibration was performed using a radionuclide source from Peking University, and neutron fluence measured by the calibrated long counter was successfully compared to values determined with reference instruments (two recoil proton counters and a 238U fission chamber). The relative deviations were lower than 6% along the whole energy range. The calibrated long counter was thus successfully applied in the calibration of a Bonner sphere spectrometer.
更多
查看译文
关键词
Long counter,Calibration,Response function,Effective center