Investigation of 3D diamond detector dosimetric characteristics

JOURNAL OF INSTRUMENTATION(2018)

引用 5|浏览66
暂无评分
摘要
Recently, a polycrystalline chemical vapor deposited (pCVD) 3D diamond detector with graphitic in bulk electrodes, fabricated using a pulsed laser technique has been evaluated for photon beam radiation dosimetry during in-air exposure. The same 3D diamond detector, has now been investigated to evaluate its performance under clinically relevant conditions putting the detector inside a Polymethylmethacrylate (PMMA) phantom, to obtain higher precision dosimetric measurements. The detector leakage current was of the order of +/- 25 pA or less for bias voltages up to -100 V. The 3D detector was tested for time stability and repeatability showing excellent performance with less than 0.6% signal variation. It also showed a linear response for low dose rates with a deviation from linearity of 2%. It was also possible to verify the detector response as a function of the depth in PMMA up to 18 cm.