Large Angle Optical Access in a Sub-Kelvin Cryostat

S. Hähnle,J. Bueno, R. Huiting, S. J. C. Yates,J. J. A. Baselmans

Journal of Low Temperature Physics(2018)

引用 5|浏览15
暂无评分
摘要
The development of lens-antenna-coupled aluminum-based microwave kinetic inductance detectors (MKIDs) and on-chip spectrometers needs a dedicated cryogenic setup to measure the beam patterns of the lens-antenna system over a large angular throughput and broad frequency range. This requires a careful design since the MKID has to be cooled to temperatures below 300 mK to operate effectively. We developed such a cryostat with a large opening angle θ = ± 37.8^∘ and an optical access with a low-pass edge at 950 GHz . The system is based upon a commercial pulse tube cooled 3 K system with a ^4He – ^3He sorption cooler to allow base temperatures below 300 mK . A careful study of the spectral and geometric throughput was performed to minimize thermal loading on the cold stage, allowing a base temperature of 265 mK . Radio-transparent multi-layer-insulation was employed as a recent development in filter technology to efficiently block near-infrared radiation.
更多
查看译文
关键词
Cryogenic optics,Cryostat,Kinetic inductance detectors,Lens-antenna
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要