New read-out electronics for ICARUS-T600 liquid Argon TPC. Description, simulation and tests of the new front-end and ADC system

L. Bagby, B. Baibussinov,V. Bellini,M. Bonesini, A. Braggiotti, L. Castellani, S. Centro,T. Cervi, A. G. Cocco,F. Fabris, A. Falcone,C. Farnese,A. Fava, F. Fichera, D. Franciotti, G. Galet, D. Gibin, A. Guglielmi,R. Guida, W. Ketchum, S. Marchini,A. Menegolli,G. Meng, G. Menon,C. Montanari, M. Nessi,M. Nicoletto, R. Pedrotta,P. Picchi,F. Pietropaolo, G. Rampazzo, A. Rappoldi,G. L. Raselli,M. Rossella, C. Rubbia, A. Scaramelli, F. Sergiampietri, M. Spanu,M. Torti,F. Tortorici, F. Varanini,S. Ventura, C. Vignoli,A. Zani, P. G. Zatti

arXiv: Instrumentation and Detectors(2018)

引用 6|浏览33
暂无评分
摘要
The ICARUS T600, a liquid argon time projection chamber (LAr-TPC) detector mainly devoted to neutrino physics, underwent a major overhauling at CERN in 2016-2017, which included also a new design of the read-out electronics, in view of its operation in Fermilab on the Short Baseline Neutrino (SBN) beam from 2019. The new more compact electronics showed capability of handling more efficiently the signals also in the intermediate Induction 2 wire plane with a significant increase of signal to noise (S/N), allowing for charge measurement also in this view. The new front-end and the analog to digital conversion (ADC) system are presented together with the results of the tests on 50 liters liquid argon TPC performed at CERN with cosmic rays.
更多
查看译文
关键词
Front-end electronics for detector readout,Neutrino detectors,Noble liquid detectors (scintillation, ionization, double-phase),Time projection Chambers (TPC)