A facility for radiation hardness studies based on the Bern medical cyclotron

arXiv: Instrumentation and Detectors(2018)

引用 23|浏览8
暂无评分
摘要
The development of instrumentation to be operated in high-radiation environments is one of the main challenges in fundamental research. Besides space and nuclear applications, particle physics experiments also need radiation-hard devices. The focus of this paper is a new irradiation facility based on the medical cyclotron located at the Bern University Hospital (Insespital), which is used as a controlled 18 MeV proton source. The adjustable beam current allows for dose rate dependent characterisation over a large dynamic range, from 0.1 to 1000 Grad per hour. The beam can be tuned so that the user can obtain the desired irradiation conditions. A complete study of the device under irradiation is possible thanks to dedicated beam monitoring systems as well as a power control system for the device under irradiation, which can be operated on-line. Further characterisations of the irradiated devices are possible thanks to a laboratory equipped with gamma spectroscopy detectors, ammeters and transient current technique setups.