A Compact 3t All Hts Cryogen-Free Mri System

B. J. Parkinson, R. A. Slade, K. Bouloukakis

SUPERCONDUCTOR SCIENCE & TECHNOLOGY(2017)

引用 30|浏览6
暂无评分
摘要
We have designed and built a passively shielded, cryogen-free 3 T 160 mm bore bismuth strontium calcium copper oxide HTS magnet with shielded gradient coils suitable for use in small animal imaging applications. The magnet is cooled to approximately 16 K using a two-stage cryocooler and is operated at 200 A. The magnet has been passively shimmed so as to achieve +/- 10 parts per million (ppm) homogeneity over a 60 mm diameter imaging volume. We have demonstrated that B-0 temporal stability is fit-for-purpose despite the magnet operating in the driven mode. The system has produced good quality spin-echo and gradient echo images. This compact HTS-MRI system is emerging as a true alternative to conventional low temperature superconductor based cryogen-free MRI systems, with much more efficient cryogenics since it operates entirely from a single phase alternating current electrical supply.
更多
查看译文
关键词
cryogen-free, HTS, MRI